Microplate ELISA Readers

For Quality Control of our Microplate Readers please check our "QC Tools" section !

LMR 340M

is a MONOCHROMATOR BASED Microplate Reader that incorporates a Fastie-Ebert monochromator with a holographic grating for easy wavelength selection without the need for filters. Any wavelength from 340 to 800 nm can be selected. The Reader is also very flexible when it comes to plate formats, taking advantage of the fact that this is a single channel instrument without the restrictions imposed by multi-channel systems. There can be total confi dence in the results too, since the LMR 340M calibrates and checks itself regularly and automatically..

Features
  • Monochromator instrument for total wavelength flexibility without the needs for filters
  • Full wavelength scan of a single well in less than 5 seconds
  • Built-in wavelength calibration
  • Can measure any plate format from 12 to 96, or 96 and 384 wells (HD version)
  • Fast reading, 20 seconds only for a 96 well plate
  • Low profile, self-centring plate carrier and front loading allows easy integration into automated systems

LMR SA

is a Stand Alone Microplate Reader that combines a large graphics display with an easy to use, yet powerful on-board software, making it ideal for use in any laboratory performing ELISA and other microplate assays. It can perform single- and dual wavelength endpoint and kinetics measurements.

Features Benefits
  • Digital light control system:
  • No re-adjustment during lifetime
  • Extra long-time stability
  • Self check and self-calibration before each measurement
  • Digital technology reduces component count and increases MTBF (Mean Time Between Failures)
  • 130 methods and 100 results can be stored
  • Sufficient for demanding laboratories.
  • Extensive range of built-in printer drivers
  • Compatibility with most printers
  • Windows® based PC-simulation program
  • Method definition, method up - and download, saving of methods and results on PC
  • QC-plate and QC-program
  • User can perform regular validation
  • Built with industrial grade components
  • Lasts for many years even under very high use

 

LMR 8

extends your evaluation possibilities beyond todays standards. Due to the multichannel design of the optical system, LMR 8 is able to read an entire 96 well plate in just 5 seconds allowing fast kinetic studies. Onboard automatic sys-checks ensure safety throughout the measurement procedures.

Key Features LMR 8:

  • Fully Computer controlled

  • Each model interfaces with several PC software packages for extended data reduction.

  • 1 Channel or 8 Channel models available

  • Each model comes with 4 Filters (free configuration on time of order)

  • Each model as UV version available

  • LMR 1 available with Grating Monochromator

  • Variable Plate sizes (6, 12, 48, 96 wells, or 96 & 384 wells)

  • Each model with robot friendly plate carrier

 

Specifications all models:

Parameter

LMR 340M LMR SA LMR 8
Measurement Range: 0.000 - 3.700 OD 0.000 - 4.000 OD 0.000 - 4.000 OD
Wavelength Range: 340 - 800nm LMR SA: 400 - 800nm
LMR SA UV: 340 - 800nm
LMR 8: 400 - 800nm
LMR 8 UV: 340 - 800nm
Reading Speed: 20 seconds for 96 well plate, single wavelength 5 seconds for 96 well plate, single wavelength 5 seconds for 96 well plate, single wavelength
Filters: Wavelength selection Monochromator; all wavelengths are selectable in 1 nm intervals, Bandwidth < 3nm 4 Filters onboard, up to 6 possible
Standard Version: 405, 450, 492, 620,
UV Version: 340, 405, 492, 620nm
4 Filters onboard, up to 6 possible
Standard Version: 405, 450, 492, 620,
UV Version: 340, 405, 492, 620nm
Plate Types: 12 - 96 wells (standard) or 96-384 wells (HD version) 96 wells 96 wells
Accuracy: +/- 1% (up to 2.0 OD) +/- 1% and 0.005 OD at 2.5 OD +/- 1% (up to 2.0 OD)
Precision: +/- 0.5% (up to 2.5 OD) +/- 0.5% and 0.005 OD at 2 OD +/- 0.5% (up to 2.5 OD)
Light Source: Quartz Halogen Lamp 30W Quartz Halogen Lamp 50W
Expected life time: 10 Million readings !
Quartz Halogen Lamp 50W
Expected life time: 10 Million readings !
Optical System: 1-Channel system with self calibration and digital light control 8-Channel system with self calibration and digital light control 8-Channel system with self calibration and digital light control
Interface: RS 232 bidirectional, USB RS 232 bidirectional, Centronics parallel RS 232 bidirectional
Display: N.A. Graphical LCD display 240 x 128 dots with backlight N.A.
Keyboard: N.A. 23-keys membrane keypad with tactile feedback N.A.
Power requirements: 90 - 250V, 50 or 60 Hz (autosensing) 90 - 250V, 50 or 60 Hz (autosensing) 90 - 250V, 50 or 60 Hz (autosensing)
Measures (W x D x H): 430 x 270 x 240 mm (W x D x H), 10kg 430 x 270 x 240 mm, 12kg 410 x 250 x 170 mm, 10kg